Objective: OXRAY is a gimbal-mounted O-ring linear accelerator that can shift the radiation isocenter away from the machine isocenter and deliver biaxial rotational dynamic radiation therapy (BROAD-RT). Published commissioning studies have characterized dynamic beam positioning with in-house analysis, but static gimbal-shifted positions have not been evaluated with a commercially available quality assurance system. We assessed a commercial Winston-Lutz (WL) and off-axis Winston-Lutz (OAWL) procedure for field-to-target coincidence at the machine isocenter and at ±30 mm along the superior-inferior axis. Approach: A MultiMet-WL Cube phantom and DoseLab Pro were used to analyze cine electronic portal imaging device images acquired during static four-port irradiation, 360° gantry rotations, and a BROAD-RT trajectory. Main results: Static 2D displacement was 0.38 ± 0.20 mm (maximum 0.73 mm) at the machine isocenter, 0.45 ± 0.15 mm (0.70 mm) at +30 mm, and 0.45 ± 0.20 mm (0.79 mm) at -30 mm. During gantry rotation, mean displacements were 0.25-0.34 mm and the maximum was 0.68 mm. BROAD-RT yielded 0.32 ± 0.16 mm with a maximum of 0.65 mm. Clockwise and counterclockwise rotations showed similar angle-dependent profiles within the measurement session. The maximum recorded-minus-commanded gimbal tilt residual was 0.004°, showing high agreement within the control system. All measured values were below 1 mm. Significance: This commercial procedure directly tests static gimbal-shifted positions that are not covered by center-only WL testing and provides the same angle-resolved endpoint for rotational and BROAD-RT delivery without custom phantom hardware or analysis software.
No takes yet. Share an insight, caveat, or question.
Aoyama et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: