Ingestible antennas have proven their usefulness in diagnosing the gastro intestinal (GI) tract, but they face challenges such as miniaturization, omni-directionality, and specific absorption rate (SAR) in different tissues with varying electromagnetic properties. This work presents an ultra-miniaturized ingestible antenna (with a footprint of 5 mm × 4.2 mm × 0.12 mm) for a wireless capsule endoscopic system. The designed antenna system supports medical device radio communications service (MedRadio) (401-406 MHz), medical implant communication systems (MICS) (402-405 MHz), industrial, scientific and medical (ISM) (433.1-434.8 MHz), and mid-field 1.2 GHz (1.24-1.34 GHz) and 2.2 GHz (2.1-2.3 GHz) bands. The SAR values in various implanted tissues have been calculated, which are in accordance with the standard limits, making the antenna safe for use in the human body. The maximum SAR values at −16 dBm are 0.01242 W/kg and 0.00209 W/kg for 1 g and 10 g tissue averaging, respectively. The corresponding maximum allowable input powers are approximately 3.2 mW and 26.5 mW, respectively. The presented antenna is encased in a capsule and tested in a heterogeneous phantom. The capsule size is reduced to 17 mm × 8 mm (typically 26 mm × 11 mm), making it one of the smallest capsules for planar multiband antennas suitable for biomedical applications.