PURPOSE: To quantitatively evaluate the optical effects of the Image Guidance System filter on surgical microscope illumination under different light source conditions. SETTING: Conducted in a darkroom laboratory environment using a surgical microscope at [blinded]. DESIGN: Experimental optical bench study under controlled laboratory conditions. METHODS: Optical parameters were measured using a surgical microscope with and without an Image Guidance System under light-emitting diode (LED), xenon, and halogen light sources. Illuminance (lux) and color temperature (Kelvin) were measured using a spectroradiometer. Spectral attenuation rate was analyzed across the visible spectrum (380-780 nm). Resolution was evaluated using a United States Air Force 1951 resolution test target. Contrast was calculated using Michelson contrast values. RESULTS: Illuminance was significantly lower in the Guidance group across all light sources (LED: 44.84 ± 0.11 vs. 27.42 ± 0.16 lux; xenon: 32.06 ± 1.01 vs. 18.70 ± 0.34 lux; halogen: 41.76 ± 0.66 vs. 23.84 ± 0.43 lux; P = .01, P = .01, P = .01). The filter reduced illuminance by 38.8%-42.9%. Spectral attenuation ranged from 35% to 50%, with an inverted U-shaped pattern. Resolution decreased significantly by 28.1% and contrast by 52.9% (P < .01). CONCLUSIONS: The Image Guidance System filter reduces illuminance by 40%, resolution by 28.1%, and contrast by 52.9%. Increasing illumination to compensate may raise the risk of retinal phototoxicity.