Key result
Optimized loss-function HRV model achieves ~86% accuracy for sleep staging against PSG.
Why the study?
Many consumer wearables are criticized for giving inaccurate sleep feedback that can lead to negative daytime consequences, motivating the optimization of heart-rate variability-based sleep classification to minimize erroneous classification during ambulatory sleep sensing.
Observational (n=136)
No
Answers are grounded in this paper and cite their sources.
Supports wearable HRV sleep staging in poor sleepers; leaves open clinical adoption pending prospective validation.
More and more people quantify their sleep using wearables and are becoming obsessed in their pursuit of optimal sleep ("orthosomnia"). However, it is criticized that many of these wearables are giving inaccurate feedback and can even lead to negative daytime consequences. Acknowledging these facts, we here optimize our previously suggested sleep classification procedure in a new sample of 136 self-reported poor sleepers to minimize erroneous classification during ambulatory sleep sensing. Firstly, we introduce an advanced interbeat-interval (IBI) quality control using a random forest method to account for wearable recordings in naturalistic and more noisy settings. We further aim to improve sleep classification by opting for a loss function model instead of the overall epoch-by-epoch accuracy to avoid model biases towards the majority class (i.e., "light sleep"). Using these implementations, we compare the classification performance between the optimized (loss function model) and the accuracy model. We use signals derived from PSG, one-channel ECG, and two consumer wearables: the ECG breast belt Polar® H10 (H10) and the Polar® Verity Sense (VS), an optical Photoplethysmography (PPG) heart-rate sensor. The results reveal a high overall accuracy for the loss function in ECG (86.3 %, κ = 0.79), as well as the H10 (84.4%, κ = 0.76), and VS (84.2%, κ = 0.75) sensors, with improvements in deep sleep and wake. In addition, the new optimized model displays moderate to high correlations and agreement with PSG on primary sleep parameters, while measures of reliability, expressed in intra-class correlations, suggest excellent reliability for most sleep parameters. Finally, it is demonstrated that the new model is still classifying sleep accurately in 4-classes in users taking heart-affecting and/or psychoactive medication, which can be considered a prerequisite in older individuals with or without common disorders. Further improving and validating automatic sleep stage classification algorithms based on signals from affordable wearables may resolve existing scepticism and open the door for such approaches in clinical practice.
No takes yet. Share an insight, caveat, or question.
Topalidis et al. (2023) conducted an observational in Self-reported sleep complaints / poor sleepers (n=136). Optimized loss-function sleep classification model with IBI quality control (using ECG, Polar H10, and Polar Verity Sense) vs. Accuracy-optimized sleep classification model and gold-standard Polysomnography (PSG) was evaluated on 4-class sleep stage classification accuracy and Cohen's kappa compared to gold-standard PSG (ECG sensor) (p=<0.001). The optimized loss-function HRV sleep classification model achieved high 4-class accuracy against PSG for ECG, Polar H10, and Verity Sense.
Synapse has enriched 10 related papers. Consider them for comparative context: