Answers are grounded in this paper and cite their sources.
The phonocardiogram (PCG) is capable of detecting structural cardiac anomalies, enabling real-time wearable monitoring using integrated neural networks (NNs). However, it encounters challenges such as fluctuations in diagnostic accuracy due to intra- and inter-individual variability, underutilization of NN hardware across heterogeneous algorithms, and unnecessary energy loss caused by always-on analog-front-ends (AFEs) and convolutional NNs (CNNs). This article presents a fully integrated, robust, and extremely low-power PCG monitoring System-on-Chip (SoC) that utilizes CNNs. It features three key techniques: 1) a dynamic framing mechanism combined with on-chip training to enhance classification stability against individual variability; 2) a reconfigurable computing array (RCA) that supports filtering, dynamic time warping (DTW), inference, and training to improve hardware utilization; 3) a comprehensive clock- and power-gating scheme for the AFE and CNN to boost energy efficiency. The prototype PCG SoC, using a 28-nm CMOS process, achieves 98.9% classification accuracy on the Yaseen dataset. In addition, it maintains an average accuracy of 97.1% with a standard deviation of 0.72% across the evaluated heart-rate range of 60-120 bpm, while consuming only 0.36 μJ per classification.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: