In the present work, a new all-solid-state microsensor built on a polyvinyl chloride (PVC) membrane was fabricated to quantify paroxetine (PRX) in pharmaceutical formulations through both potentiometric and voltammetric approaches. An ion-pair complex between paroxetine and tetraphenylborate (PRX–TPB) was synthesized and incorporated as the ionophore within the PVC membrane. Detailed potentiometric characterization showed that the sensor followed Nernstian behavior, yielding a slope of 59.1 ± 0.8 mV per decade (R 2 = 0.9998) across paroxetine hydrochloride concentrations spanning 10 –6 to 10 –2 mol.L -1 . The microsensor reached a detection limit of 3.71 × 10 –7 mol.L −1 , combined with a response time below 10 s and roughly seven weeks of usable working life without loss of potentiometric behavior. Best performance fell within the pH window of 4.0 to 8.0. The same PRX-selective device was further characterized voltammetrically and successfully applied to paroxetine assay in marketed tablets by both electrochemical routes, with outcomes matching UV–Vis spectrophotometry at 95% confidence level, providing evidence that the microsensor is a dependable, uncomplicated, and low-cost option for pharmaceutical quality control.