Catecholamines, including dopamine, epinephrine, and norepinephrine, are important biomarkers of neurological and neuroendocrine disorders. However, their simultaneous quantification in body fluids remains challenging because of their similar molecular structures, which produce highly overlapping Raman spectroscopic signatures. Here, we develop an integrated platform combining an iron-nitrilotriacetic acid (NTA-Fe)-functionalized surface-enhanced Raman scattering (SERS) paper with a two-stage machine learning pipeline for catecholamine quantification in artificial urine. The SERS paper enables rapid drop-and-measure detection, achieving limits of detection of 7.18, 9.01, and 6.80 μM for dopamine, epinephrine, and norepinephrine in artificial urine, respectively. For samples with unknown compositions, the pipeline first classifies the catecholamines present (92.19% group-level accuracy) and then quantifies single-component samples via intensity-ratio calibration and mixtures via trained partial least squares regression models (global root-mean-square error: 2.27 μM). This work establishes a functionalized SERS paper platform coupled with machine learning as a proof of concept for the identification and quantitative analysis of structurally similar metabolites in complex mixtures.