Answers are grounded in this paper and cite their sources.
Real-time dose-guided multileaf collimator tracking uses accumulated dose estimates to adapt subsequent irradiation during target motion. This Mini Review examines when dose history may improve dose delivery beyond geometric or fluence-based tracking. Five representative studies are compared across single-target motion, differential multi-target motion, and anatomical deformation. Single-target simulations showed improved agreement between modeled and planned dose with dose optimization, whereas a phantom experiment on a magnetic resonance imaging–guided linear accelerator (MRI-linac) yielded numerically similar results for dose-guided and geometric tracking. In a selected dual-target geometry with overlapping beam projections, dose-guided tracking improved pooled measured dose agreement, with benefit dependent on the motion scenario. Deformation studies demonstrated the importance of tissue correspondence, but comparisons retaining dose history in both arms primarily tested motion representation. The available evidence suggests that history may be useful when accumulated dose errors alter aperture decisions and the remaining irradiation allows those errors to be corrected. Feedback on actual delivery and organ-at-risk objectives may help translate modeled gains into delivered-dose improvements. Comparisons that explicitly isolate accumulated-dose feedback, followed by independent dosimetric verification, are needed to establish where dose history provides an incremental advantage.
No takes yet. Share an insight, caveat, or question.
Shichao Zhang (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: