Purpose To establish an in vitro porcine ureteral stone model, identify factors affecting ureteral inner-wall thermal injury during holmium laser lithotripsy, and construct a predictive model. Methods Fresh porcine urinary systems were used. A human stone (7–10 × 5–7 mm) was implanted into the upper ureter. A temperature sensor was fixed on the inner wall below the stone, and a 365 µm holmium laser fiber was placed below the stone. Perfusion was 20 or 40 mL/min; pulse width was long or short; power was 10 or 20 W (four modes). Activation lasted 10 or 20 s; each condition was repeated 3 times. Results Ninety-six measurements were obtained (53.18 ± 24.71°C; range 27.28–105.10°C). Perfusion rate was strongly negatively correlated with temperature (r = − 0.905, p < 0.001); power was moderately positively correlated (r = 0.210, p = 0.020); pulse width showed no significant correlation (r = 0.056, p = 0.295). The model included time, perfusion rate, power, and pulse width (R²= 0.874; F(4,91) = 157.407, p < 0.001). At 20 mL/min, maximum temperatures exceeded 43°C; at 40 mL/min, all remained below 43°C. Conclusion Perfusion rate was the strongest negative factor. Increasing perfusion from 20 to 40 mL/min reduced temperature by 44.5°C; each 10 W increase in power raised it by 10.3°C; doubling activation time raised it by only 4.3°C. Pulse width had no significant effect. Low power (10 W) is recommended; 40 mL/min maintained temperatures below 43°C.