Quantitative measurements of piezoelectric response using piezoresponse force microscopy (PFM) are often compromised by non‐piezoelectric contributions, particularly electrostatic interactions. Interferometric displacement sensor (IDS)‐based PFM can directly measure tip displacement and thereby suppress many of these artifacts. However, accurate determination of the piezoelectric response remains highly sensitive to measurement conditions. Here, the combined effects of laser‐spot position, surface‐potential compensation, and cantilever stiffness on the measured vertical effective piezoelectric coefficient d eff are systematically evaluated. Precise on‐tip alignment of the IDS laser spot is shown to be a prerequisite for capturing the true tip displacement. Only when this condition is met and the local surface potential is properly compensated, does the piezoresponse amplitude between oppositely polarized domains converge across all tested conditions. Soft cantilevers are more susceptible to electrostatic artifacts than stiff cantilevers. A three‐step protocol for quantitative IDS‐PFM measurements is established, providing a framework for the metrology of emerging electromechanical materials.