Why the study?
High-density electroanatomical mapping is routinely used for scar-related VT ablation, but the performance of multielectrode mapping catheters with different array designs had not been directly compared.
Population
104 consecutive patients undergoing left ventricular scar-related VT ablation
Comparison
Flexible spline catheter (n = 44) vs fixed paddle catheter (n = 60)
Design
Retrospective analysis
Answers are grounded in this paper and cite their sources.
Comparable acute success with spline versus paddle catheters; leaves open whether design affects VT recurrence or mapping in scar-related ablation.
BACKGROUND: High-density (HD) electroanatomical mapping is routinely used for scar-related ventricular tachycardia (VT) ablation. The performance of multielectrode mapping catheters with different array designs has not been directly compared. OBJECTIVE: To compare the procedural performance of two different multielectrode mapping catheter designs-a flexible spline catheter (SC) and a fixed paddle catheter (PC) in scar-related VT ablation. METHODS: We retrospectively analyzed 104 consecutive patients who underwent left ventricular scar-related VT ablation (SC, n = 44; PC, n = 60). The primary endpoint was acute procedural success, defined as non-inducibility of the clinical VT after ablation. Secondary endpoints included procedural efficiency, VT inducibility, VT morphology burden, and procedural and in-hospital adverse events. RESULTS: Acute procedural success was achieved in 95.5% of SC and 98.3% of PC procedures (p = 0.57). Procedure duration, LV mapping time, mapping point count, fluoroscopy time, and adverse-event rates were comparable between groups. Overall VT inducibility (mechanical or programmed induction) was similar (81.8% vs. 76.7%; p = 0.63); however, ≥3 distinct inducible VT morphologies were significantly more frequent with the spline catheter than with the paddle catheter (38.6% vs. 11.7%; p = 0.002). During follow-up, rates of VT recurrence (34.5% vs. 20.0%; p = 0.19) and repeat ablation (13.8% vs. 7.3%; p = 0.44) did not differ significantly between groups. CONCLUSIONS: Both catheter designs achieved comparable acute procedural success, procedural efficiency, and safety in scar-related VT ablation. Procedures with the spline catheter were associated with a higher inducible VT morphology burden; this hypothesis-generating finding, which may partly reflect differences in substrate location, warrants prospective evaluation.
No takes yet. Share an insight, caveat, or question.
Trajkovska et al. (2026) studied this question.