Conventional Gaussian-beam second-harmonic generation (SHG) microscopy requires time-consuming axial scanning to image across tissue depth. Here, we integrate a two-axicon Bessel-beam shaper into a stage-scanning SHG microscope and compare imaging across 2° and 5° axicon pairs. Characterization of SHG from a quartz slab gives depths of focus (DOFs) that are 2.37 and 1.56 times the Gaussian DOF, with continuous tuning via inter-axicon translation. Evaluation across tendon and zebrafish specimens quantifies structural visibility and contrast. Bessel-beam performance was further tested on week-10 mouse mammary glands, where a single Bessel image captures collagen structures also visible in a z-projection of 15 Gaussian planes (∼28 µm span). Relative to this protocol, one Bessel acquisition results in an estimated 15-fold speedup in acquisition time. Additional equal-input-power quartz measurements and temporal image analyses quantify the signal-to-noise trade-offs associated with extended-focus imaging.