Answers are grounded in this paper and cite their sources.
MRI residual geometric distortion compromises stereotactic body radiotherapy accuracy, but distortion magnitude, directional dependence, and bandwidth (BW) effects across individual units remain insufficiently understood. To evaluate the magnitude, spatial distribution, and directional characteristics of residual distortion across five clinical MRI units under multiple receiver-BW settings for unit-level quality assurance (QA). A 3D phantom was scanned on five 1.5T or 3T units (three vendors) at BWs of 200–1000 Hz/pixel with vendor 3D correction enabled. Absolute residual distortion and its x-, y-, and z-components were evaluated relative to isocenter. Whole-volume distortion generally decreased as BW increased, but was non-monotonic in central regions. Distortion increased with distance from isocenter, with dominant displacement axes (x-axis for Skyra, z-axis for Optima) differing among units. Residual distortion varies among individual units and is not fully characterized by BW alone, supporting unit-level geometric QA.
No takes yet. Share an insight, caveat, or question.
Washio et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: