Conventional handheld ultrasound provides localized, single-sided views of limb musculoskeletal structures. Examination quality is highly operator-dependent because probe position, beam incidence angle, acoustic-window selection, and contact pressure vary with technique, reducing spatial reproducibility. Moreover, acoustic heterogeneity causes sound-speed mismatch and focusing errors. To address these limitations, we developed a fixed ring-array ultrasound computed tomography (USCT) system for complete cross-sectional imaging and three-dimensional reconstruction. It integrates a 512-element ring array, 128-channel electronics, and mechanical axial scanning. A dataset-specific global effective sound speed was estimated by robustly fitting Tx-Rx direct-wave distances to arrival times and applied to synthetic-aperture delay-and-sum reconstruction. We evaluated a geometric phantom, two bone-soft-tissue phantoms, and two human forearms. At the estimated speed of 1507.2 m/s, mean contrast-to-noise ratio (CNR) increased by 8.9% and 11.0%, and geometric bias decreased by 54.5% and 65.1% relative to 1480 and 1540 m/s, respectively. The system visualized bone-mimicking interfaces and produced complete in vivo forearm cross sections, demonstrating the feasibility of reproducible full-circumference musculoskeletal USCT.