Answers are grounded in this paper and cite their sources.
Hydrogen peroxide (H2O2) is one of the most extensively studied reactive oxygen species (ROS), serving as a key redox signaling molecule involved in diverse redox-regulated cellular processes. For scalable electrochemical monitoring of extracellular H2O2, a submicromolar limit of detection (LOD) is required but remains challenging because background redox currents originating from the culture medium obscure the H2O2-dependent current response. Herein, we report an electrochemical H2O2 monitoring technique using a PtNP-modified indium tin oxide (ITO) electrode operated by two-step potential chronoamperometry (TCA). An anodic pretreatment step forms platinum surface oxide, which is subsequently reduced during the cathodic measurement step. This modulation of the surface oxide state enhances the H2O2-dependent response, yielding an enhanced reduction current that increases linearly with H2O2 concentration. The two-step potential modulation strategy provides robust signal amplification, achieving an LOD of 0.12 μM, which is applicable to extracellular H2O2 monitoring. The TCA-operated PtNP/ITO electrode is further applied to incubation media collected from NOX4-expressing living cells, where it reliably resolves inhibitor-dependent modulation of extracellular H2O2 levels. This work demonstrates the potential of TCA-operated PtNP/ITO electrodes for measuring extracellular H2O2 and evaluating pharmacological modulation of hydrogen peroxide secretion.
No takes yet. Share an insight, caveat, or question.
Choi et al. (2026) studied this question.