Answers are grounded in this paper and cite their sources.
Environmental exposure to potentially toxic elements is a worldwide public health concern. Inductively coupled plasma mass spectrometry (ICP-MS) is preferred for trace-level human biomonitoring but is not universally accessible. Data were generated within a Mexican cohort using inductively coupled plasma optical emission spectrometry (ICP-OES) as a more accessible multielement alternative to measure As, Cd, Cr, Pb, and Hg in paired blood and urine from 187 participants, including patients with rheumatic diseases or osteoporosis and clinically healthy participants. Unexpected concentration patterns prompted this secondary appraisal of detectability, censoring, limits of detection (LODs), population comparators, and exploratory agreement with graphite furnace atomic absorption spectrometry (GFAAS) for blood Pb. Numerical reporting varied across elements and matrices and LODs exceeded concentrations reported in population biomonitoring, making comparator exceedance analytically inevitable. Contrasting blood–urine patterns and discordance between ICP-OES and GFAAS Pb measurements raised concerns. The evaluated workflow did not demonstrate adequate fit-for-purpose performance for trace-level biomonitoring. Nevertheless, these findings document challenges for laboratories seeking accessible alternatives to ICP-MS and highlight the need for matrix-specific validation, appropriate LODs, interference assessment, and reference-method confirmation before cohort-wide implementation.
No takes yet. Share an insight, caveat, or question.
Becerra-Jimenez et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: