Answers are grounded in this paper and cite their sources.
A one-dimensional photonic crystal with a single defect [(GaP/SiO2)5/ sensing medium/ (SiO2 /GaP)5 / glass] is proposed as an optical biosensor to detect urea and albumin in urine in the near infrared region. The physical parameters of the proposed structure are optimised to attain high sensitivity. The design is numerically investigated using Transfer Matrix Method and Finite Element method. Incorporating polarisation dependent anisotropy behaviour of the analytes have helped in analyte discrimination making it suitable for multi-analyte detection. Machine learning assisted quantitative analysis for concentration of the analytes are highly consistent with values from optical characteristics of the sensor. The proposed design of biosensor exhibited a sensitivity of 915nm/RIU for urea detection and 1000nm/RIU for albumin detection with low limit of detection. The prediction metrics such as root mean square error (RMSE) and mean absolute error (MAE) of urea detection is 0.5gm/dl and 0.315gm/dl and that of albumin is 0.05 gm/dl, and 0.031gm/dl respectively. These findings indicate the potential of the proposed design for urea and albumin detection with great accuracy.
No takes yet. Share an insight, caveat, or question.
Mammen et al. (2026) studied this question.