Answers are grounded in this paper and cite their sources.
Hyaluronidase (HAase) is a clinically relevant biomarker closely associated with bladder cancer and inflammation, yet its rapid and decentralized quantification remains challenging owing to the reliance on costly instrumentation and complex signal labeling. Herein, we report a time-readout capillary sensor (TR-CapSensor) that enables portable, label-free, and visual detection of HAase and inhibitor screening. The TR-CapSensor leverages the HAase-catalyzed degradation of hyaluronic acid to trigger a substantial decrease in solution viscosity, which is physically transduced into the falling time of a tracer bead within a capillary. A smartphone records the descent process, and a custom-developed Android application converts the manually entered falling time into a digitized concentration readout. The sensor exhibits a linear response to HAase in the range of 0-10 U/mL (R2 = 0.993), with a limit of detection (LOD) of 0.98 U/mL. It further demonstrates good selectivity and anti-interference capability, achieving spiked recoveries of 86-113% in urine samples. For drug discovery, the TR-CapSensor successfully identified tannic acid as a potent HAase inhibitor, with an IC50 of 20.4 μg/mL within 35 min. This portable sensor offers a robust, cost-effective, and universally accessible tool for proof-of-concept HAase detection and inhibitor screening, with potential for future clinical translation, particularly in resource-limited settings.