Tribocorrosion of passive implant alloys is usually quantified after the fact, by splitting a measured material loss into pure wear, pure corrosion and their synergy following ASTM G119. This bookkeeping does not tell an implant designer how the synergy will change with load, sliding speed, electrode potential, protein content or surface treatment, nor how uncertain that prediction is. We present TriboSyn, a physics-based framework that couples Hertzian contact mechanics with wear-induced track widening, Archard wear with Langmuir protein boundary lubrication, asperity-swept depassivation with saturation, exponential repassivation kinetics with a potential-dependent film charge, Faraday conversion, third-body enhancement of mechanical wear and a galvanic mixed-potential model of the open-circuit potential (OCP). The framework is solved by an adaptive ordinary-differential-equation (ODE) integrator and by a vectorized closed form that reproduces the ODE solution with a mean absolute error of 1.37%; a pass-resolved simulation of the moving contact confirms the averaged current within 0.29% at physiological sliding frequency. Applied to Ti-6Al-4V, CoCrMo and 316L stainless steel sliding against alumina in phosphate-buffered saline (10 N, 1 Hz, 1 h, +0.2 V vs Ag/AgCl), the model predicts total losses of 54.9, 14.7 and 72.1 × 10−3 mm3 with synergy fractions of 42.2%, 78.2% and 77.3%, and OCP drops of 572, 229 and 191 mV. Potential–velocity mechanism maps show that the corrosion–wear boundary (Ψ = 1) lies at +0.51 V for Ti-6Al-4V but at −0.26 and −0.18 V for CoCrMo and 316L. Variance-based Sobol analysis attributes 54% of the variance in dissolved metal to the repassivation pair (ipk, τ ), and Monte Carlo propagation gives 90% intervals that span roughly one order of magnitude. A surface-engineering design map and an OCP-coupled fretting model of a ceramic-head modular taper show that metal release at the taper is limited by cathodic capacity rather than by micromotion amplitude. The framework is intended for calibration against new experiments.
No takes yet. Share an insight, caveat, or question.
M et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: