In vitro electrophysiological recordings of cardiac action potentials are crucial for the assessment of cardiac toxicity in drug development and cardiovascular research. Organic electrochemical transistors (OECTs) have recently emerged as promising candidates for multifunctional in vitro biosensors due to their sustainable material composition, cost‐effective fabrication processes, and multifunctional capabilities for sensing different physicochemical quantities in liquid. In this study, we introduce a novel method for recording intracellular cardiac action potentials by integrating planar OECTs with laser optoporation. Laser poration induces transient nanopores in the cell membrane, enabling direct and efficient electrical coupling between the electrodes and the intracellular compartment. No adverse effects are observed, and long‐term monitoring can be performed over several days. This approach leverages the simplicity and efficiency of planar OECT fabrication alongside the precision of laser optoporation to achieve high‐resolution intracellular recordings. Our results demonstrate the potential of this combined technique to advance cardiac toxicity assessment, offering a highly sensitive, sustainable, and scalable solution for electrophysiological studies.