Answers are grounded in this paper and cite their sources.
Background Anterior cervical corpectomy and fusion (ACCF) is an essential surgical procedure for cervical spondylotic myelopathy and other cervical pathologies. Conventional titanium mesh cage (TTMC) combined with anterior plate fixation provides immediate stability but is associated with postoperative dysphagia, subsidence, and screw loosening. To address these drawbacks, a novel 3D-printed self-stabilizing titanium mesh cage (3D-STMC) was designed with patient-specific anti-subsidence rings and integrated self-stabilizing fixation plates for zero-profile reconstruction. Its biomechanical performance was evaluated via finite element (FE) analysis and in vitro ovine cervical testing. Methods A C2–C7 three-dimensional finite element model was reconstructed from cervical computed tomography data. Two reconstruction strategies following C5 corpectomy were simulated. The first employed TTMC supplemented with an anterior cervical plate. The second employed standalone implantation of the 3D-STMC. The range of motion (ROM) of the operative segment was quantified. Peak von Mises stresses at the mesh-endplate interface and screw-bone interface were also measured. Adjacent-segment disc stresses were evaluated. All parameters were assessed under flexion, extension, lateral bending, and axial rotation. In parallel, fresh ovine cervical specimens were used to establish C5 corpectomy models. These were instrumented with either the 3D-STMC or the TTMC-plate construct. Results Finite element analysis showed operative-segment range of motion reductions exceeding 96.2% (3D-STMC) and 94.1% (TTMC) versus the Intact. The 3D-STMC reduced mesh-endplate peak stress by >20% and screw-bone interfacial stress by >43.9% versus TTMC. Adjacent-segment disc stress varied by <10% in both constructs. Ovine testing confirmed ROM reductions of >72.3% (3D-STMC) and >70.7% (TTMC). No significant intergroup differences were detected (P > 0.05). Conclusion Without anterior plate augmentation, the 3D-STMC provides equivalent immediate biomechanical stability to the conventional TTMC-plate construct. It substantially mitigates stress concentrations at the mesh-endplate and screw-bone interfaces without increasing mechanical burdens at adjacent segments. Its zero-profile design potentially reduces plate-related dysphagia. These findings indicate promising prospects for clinical translation.
No takes yet. Share an insight, caveat, or question.
龚灏 et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: