Answers are grounded in this paper and cite their sources.
In this paper, metamaterial absorber (MTM-A) for terahertz (THz)-based biomedical sensing is presented. The proposed assembly involves a symmetrically spotted metallic resonator, a SiO 2 dielectric base and a continuous metallic ground plane to achieve solid electromagnetic confinement and efficient absorption. Numerical analysis shows a maximum absorption of 99.95% at 4.52 THz, and a minimum reflection coefficient of 0.015. The symmetric unit-cell configuration retains a consistent resonant response for different polarization orientations. Sensing performance is investigated by changing the refractive index of a 1.5-um-thick analyte layer from 1.35 to 1.41. The resulting resonance shift from 4.55 to 4.42 THz confirms the structure capability to sense minor variations in dielectric characteristics of biological samples. The sensor exhibits a sensitivity of ~ 0.55 THz/RIU, Q-factor ~ 10.5 and figure of merit (FOM) of 1.55 RIU⁻¹. Besides, numerical simulations show distinct spectral characteristics for healthy and diseased biological specimens such as cancer-cell types. These results demonstrate the promise of the presented absorber as a compact, label-free THz sensing platform for biomedical analysis.
No takes yet. Share an insight, caveat, or question.
Reddy et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: