Answers are grounded in this paper and cite their sources.
Background Although frequently deemed a subtle technical detail, polyethylene liner design fundamentally governs construct stability, biomechanics, and wear. Methods This review synthesizes current evidence on liner constraint metrics, joint kinematics, and wear dynamics. Results Significant variability exists in liner capture across manufacturers, with “standard” liners ranging from 35% to 63% and “retentive” liners ranging from 45% to 81% capture. The liner stability ratio indicates that 135° systems require a higher constraint (liner stability ratio [LSR] > 184%) than 155° systems (LSR 147–152%) to maintain equivalent stability. Liner thickness affects both lateralization and distalization simultaneously, though the liner contributes far less to total lateralization than the glenoid or humeral components. Highly cross-linked polyethylene shows approximately 56% less wear than conventional ultra-high-molecular-weight polyethylene, and vitamin E-enhanced polyethylene shows a 36–49% reduction in simulator studies. Constrained liners improve stability directionally, and the 26° motion cost measured in simulation is not reproduced in matched clinical cohorts. Asymmetric liner wear accounts for 60% of late dislocations. Discussion To optimize outcomes in reverse total shoulder arthroplasty, polyethylene liner constraint must be intentionally paired with humeral stem geometry rather than guided by manufacturer classification.
No takes yet. Share an insight, caveat, or question.
Kress et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: