Answers are grounded in this paper and cite their sources.
Saline immersion alters the electrical environment of the third-space tunnel, yet guidance on the electrosurgical mode selection under saline is lacking. We describe a saline-adapted two-program generator workflow for esophageal underwater third-space endoscopy in a single-center, single-operator series of 40 consecutive patients analyzed retrospectively. An arc-dependent cutting mode, Endo Cut I (manufacturer nominal peak voltage [Upmax] 550 Vp), was retained for all tissue divisions, while the arc-dependent coagulation mode Forced Coag (Upmax 1430 Vp) was replaced by the low-voltage, non-arcing Soft Coag (Upmax 110 Vp). Soft Coag was the sole coagulation mode in 38 of the 40 procedures; Forced Coag was used in two early cases, achieved rapid hemostasis in both, and was abandoned because its activation produced visible sparking and coalescent bubbles that obscured the field. Technical success was 40/40 (100%; 95% CI, 91.2–100%). One patient, within the first five cases, required coagulation forceps during myotomy at the esophagogastric junction and developed pneumoperitoneum treated by needle decompression (AGREE IIIa); no other adverse event occurred. The bubble observation is unblinded, the learning curve is confounded with the protocol change, and no superiority is claimed.
No takes yet. Share an insight, caveat, or question.
Do et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: