Key result
Very high-power short-duration PVI triggers significant esophageal temperature elevations in ~38% of patients.
Why the study?
Does pulmonary vein isolation using very high-power short-duration (90 W/4 second) radiofrequency ablation at the posterior wall cause luminal esophageal temperature elevations ≥ 39°C in patients with atrial fibrillation?
Observational (n=107)
Does pulmonary vein isolation using very high-power short-duration (90 W/4 second) radiofrequency ablation at the posterior wall cause luminal esophageal temperature elevations ≥ 39°C in patients with atrial fibrillation?
Very high-power short-duration (90 W/4 second) ablation at the posterior wall frequently causes delayed luminal esophageal temperature elevations ≥ 39°C with prolonged thermal decay, particularly near the left inferior pulmonary vein, warranting temperature monitoring and avoidance of rapid adjacent applications.
Answers are grounded in this paper and cite their sources.
May warrant esophageal temperature monitoring during very high-power short-duration PVI; leaves open the risk of clinical injury in prospective studies.
Very high-power short-duration (VHPSD) ablation has been proposed to produce shallower lesions due to a predominance of resistive over conductive heating, potentially reducing collateral tissue injury. However, in vivo data on luminal esophageal temperature (LET) during VHPSD pulmonary vein isolation (PVI) remain limited. A total of 107 consecutive patients (62 ± 10 years; 75 male; 88 with paroxysmal atrial fibrillation) underwent PVI using VHPSD (90 W/4 second) at the posterior wall and high-power short-duration (HPSD; 50 W, ablation index 450) at the anterior wall. LET was continuously monitored during the procedure, and all sites with LET elevation ≥ 39°C were analyzed. LET ≥ 39°C occurred in 41 patients (38.3%), with a mean peak LET of 40.05 ± 0.99°C. The mean time to peak LET was 13.9 seconds, while temperature returned to within 1°C of baseline after 65 seconds following cessation of radiofrequency (RF) delivery. LET elevations ≥ 39°C occurred at distances of 3.76 ± 1.73 mm in the anteroposterior (AP) view and 2.83 ± 1.10 mm in the left anterior oblique (LAO) view from the temperature sensors. Notably, 89% of sites with LET ≥ 39°C were associated with the left inferior pulmonary vein (LIPV). VHPSD ablation is associated with LET elevations ≥ 39°C, particularly during ablation near the inferior pulmonary veins. Peak temperature occurs after completion of RF delivery, limiting real-time modulation of application duration. Given the prolonged temperature decay, rapid adjacent RF applications should be avoided. LET monitoring should be considered during VHPSD ablation.
No takes yet. Share an insight, caveat, or question.
Shin et al. (2026) conducted an observational in Atrial fibrillation (n=107). Very high-power short-duration (VHPSD; 90 W/4 s) pulmonary vein isolation at the posterior wall and HPSD (50 W) at the anterior wall was evaluated on Luminal esophageal temperature (LET) elevation ≥ 39°C. Very high-power short-duration (90 W/4 s) PVI caused luminal esophageal temperature elevations ≥39°C in 38.3% of patients, with 89% of sites near the left inferior pulmonary vein.
Synapse has enriched 10 related papers. Consider them for comparative context: