Answers are grounded in this paper and cite their sources.
Sample planarity is rarely monitored in fluorescence microscopy quality control. Here we describe FlatStat, a tool that estimates sample tilt automatically from any 3D fluorescence stack, and applied it to 1187 images from 20 public datasets. This revealed that tilt is ubiquitous across laboratories (median: 1.28 μm/100 μm), strongly structured across projects, and associated with a modest but detectable increase in chromatic shift (~57 nm). FlatStat is freely available as an ImageJ macro and Python script (doi: 10.5281/zenodo.20325392).
No takes yet. Share an insight, caveat, or question.
Jacques S. Brocard (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: