Answers are grounded in this paper and cite their sources.
Pre-eclampsia arises from early placental dysfunction that precedes clinical presentation, creating a need for molecular readouts that enable timely risk assessment. Placenta derived extracellular vesicles (EVs) carry nucleic acid biomarkers with diagnostic potential, yet current analytical workflows rely on multistep isolation and extraction procedures that are slow and resource intensive. Here, we report EV-LAMP, an extraction-free isothermal assay for detecting placental EV-derived KISS1 mRNA. Placental alkaline phosphatase (PLAP) positive choriocarcinoma-derived EVs are selectively captured from biofluids using PLAP antibody conjugated magnetic beads within 20 min. Captured EVs are treated with RNase inhibitor and lysed at 90 °C for 5 min, enabling direct RNA release and eliminating conventional extraction steps. The released RNA is analyzed in the same vessel by colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP), providing semi-quantitative readout within 40 to 60 min. The assay semi-quantitatively estimates KISS1 mRNA abundance across 103 to 108 EVs per reaction, with a regression-derived estimated limit of detection of about 2 EVs and a limit of quantification of 103 EVs. Regression-derived estimates suggest approximately one amplifiable KISS1 mRNA copy equivalent per 27 EVs, compared with one copy equivalent per 38 EVs measured by reverse transcription quantitative polymerase chain reaction (RT-qPCR). In both the EV-depleted serum spike-in and the plasma spike-in samples containing mixed vesicle populations, the assay detects target derived transcripts at 1% abundance and retains measurable signal at the quantification limit in the presence of up to 108 non-target EVs. These results establish EV-LAMP as an isothermal and simplified platform for EV-derived mRNA analysis that integrates selective capture, lysis, and amplification in a single workflow completed in 65–85 min. This method provides a general framework for semi-quantitative detection of EV-derived transcripts and can be adapted to other coding and non-coding RNA targets for broader EV-associated RNA analysis.
No takes yet. Share an insight, caveat, or question.
Shaon et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: