Answers are grounded in this paper and cite their sources.
Simultaneous monitoring of glucose and amino acids in interstitial fluid (ISF) enables a more comprehensive evaluation of metabolic dysregulation, particularly for early intervention in diabetes and related metabolic disorders. Herein, we report a multiplexed electrochemical microneedle sensor array for simultaneous detection of glucose, tyrosine (Tyr), and tryptophan (Trp). A hierarchically interconnected copper-based metal-organic framework (Cu-MOF) was designed and synthesized, which was directly grown on the resin microneedle surface via an anodic dissolution strategy, creating a high-surface-area, three-dimensional conductive scaffold. This unique framework not only provides abundant active sites for enzyme and molecularly imprinted polymer (MIP) immobilization but also benefits the rapid analyte diffusion process, achieving higher sensitivity and a longer lifetime of the microneedle-based sensor. Experimental results reveal that the proposed microneedle sensor array enables simultaneous glucose (0-18 mM), Tyr, and Trp (0-100 μM) detection. Notably, the sensor retains >85% of its initial oxidation current after 15 days of ambient storage. In Vivo and in vitro evaluations confirm clear, quantitative signal responses to physiologically relevant fluctuations of glucose, Tyr, and Trp levels. This work establishes a Cu-MOF-based multiplexed microneedle platform for minimally invasive, multi-analyte metabolic monitoring in ISF.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2026) studied this question.