A laboratory optical arrangement combines a rigid 146 deg dual-mirror target with two camera channels for subpixel laser-spot localization. A single incident beam illuminates the joined mirror region, and the two reflected paths are observed separately without a beam splitter or successive reflection of one ray by both mirrors. Spot intensity is fitted with a seven-parameter rotated elliptical Gaussian, followed by residual screening and channel-wise linear calibration. The dataset comprises 198 measurement images: three motion conditions, two camera channels, three repeats, and 11 settings per repeat. Camera 1’s best complete runs give mean absolute calibration deviations of 0.044 mm for pure translation, 0.020 deg for pure rotation, and 0.035 mm for displacement along a prescribed compound trajectory. Corresponding Camera 2 values are 0.060 mm, 0.025 deg, and 0.253 mm. Three-repeat means, sample standard deviations and standard errors are reported separately from the selected runs. Because calibration and evaluation use the same nominal setting labels, these results quantify in-sample calibration agreement rather than independent accuracy or total uncertainty. The compound analysis estimates displacement only; general independent translation–rotation separation is not established.