Answers are grounded in this paper and cite their sources.
This work presents improved timing resolution in positron emission tomography, demonstrating how depth-of-interaction and time-of-flight corrections enhance image quality.
To increase the image quality and spatial resolution in positron emission tomography (PET), organ-dedicated and large axial field of view scanners require precise correction of the parallax errors at oblique angles via depth-of-interaction methods. In addition to their large solid angle coverage, time-of-flight (TOF) information, i.e., a high coincidence time resolution (CTR), can be exploited to increase the image signal-to-noise ratio and effectively boost system sensitivity. Due to the impact of the photon travel time spread (PTS) on the TOF resolution, a high DOI resolution is essential to correct the impact of the PTS on the CTR and achieve even higher timing resolution. In this work, we present a TOF- and DOI-capable detector block with a CTR of 187 ps (FWHM) to 213 ps (FWHM), read out with the NINO and TOFPET2 ASIC, and a DOI resolution of 3.4 mm to 4.7 mm (RMSE), which distinguishes at least two DOI layers in the detector. Additionally, we established benchmarks for the CTR with custom high-frequency readout electronics. With these, a CTR of 136 ps (FWHM) can be maintained for a single TOF-DOI unit channel in the detector block.
No takes yet. Share an insight, caveat, or question.
Nadig et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: