Answers are grounded in this paper and cite their sources.
Novel surgery adjusts intraocular pressure in glaucoma patients, suggesting better outcomes and fewer complications.
The purpose of this study is to present the “Hourglass Tie” technique for Paul ® Glaucoma Implant (PGI), employing three 8/0 nylon ligatures to adjust flow during surgery and enable postoperative intraocular pressure (IOP) modifications through laser suture lysis. Two cases of advanced glaucoma were managed with PGI using this innovative approach. The first case involved a 75-year-old male with severe ocular surface disease and progressive glaucoma despite medical therapy. The second case featured a 56-year-old male with uncontrolled IOP despite maximal medications and prior glaucoma surgery. The technique involved applying an 8/0 nylon ligature at the maximum length from the limbus, followed by two proximal ligatures with the “Hourglass Tie” sign confirmed at each step. Case 1 required one suture lysis at 6 weeks, achieving stable IOP of 12 mmHg without medications at 10 months. Case 2 required two suture lyses by the 4 th week, sustaining IOP of 8 mmHg at 8 months without medications. No complications were noted. This technique ensures precise flow regulation, reduces early postoperative hypotony risk, and allows long-term IOP adjustment compared to conventional methods. The “Hourglass Tie” technique is minimally invasive, well-tolerated, and effective in achieving target IOP in adult glaucoma patients.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: