Experimental model shows CIED withstands contact-sports impacts, suggesting lower risk in certain activities.
Introduction CIED patients are routinely advised against physical activity with a risk of collision because of expected damage to the implanted device. However, no data support this practice. Methods Sixteen CIED systems (6 pacemakers, 6 ICD, 4 CRT-D) from all manufacturers were implanted subcutaneously in a porcine thorax and increasing weights were dropped on the experimental setting. CIED function was analyzed in an interactive heart simulator (InterSim III, IB Lang). In a second experiment, forces resulting in irreversible damage were investigated in explanted CIED. Results No signs of malfunction were seen in any CIED after a total of 256 impacts up to 2000 N. Forces exceeding 1500 N reproducibly led to rib fractures of the porcine thorax, in line with previous reports describing serial rib fractures in humans at similar forces. One IS1/DF1 single-coil ICD lead was bent after an impact of 900 N, however no signs of lead malfunction were noted on inspection and fluoroscopy. When testing explanted CIED to failure, only 4 of 16 CIED (25%) showed signs of damage on fluoroscopy. Two ICD were unable to terminate VF despite reporting normal shock impedances before VF induction. Conclusion This is the first study to analyze the effects of blunt-force trauma on CIED. Our data suggest that CIED systems withstand contact-sports impacts which can result in serial rib fractures in humans. Therefore, general traumatological risks could outweigh the risks of CIED damage in sports with a low risk of collision. If damage is suspected, device interrogation should be performed.
No takes yet. Share an insight, caveat, or question.
Wegner et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: