Key result
Direct irradiation of cardiac implantable electronic devices up to 20 Gy caused no malfunctions when shock therapy was deactivated, suggesting total body irradiation is feasible and safe.
Why the study?
Direct irradiation of CIEDs has traditionally been avoided, leaving the feasibility of total body irradiation with a CIED unaddressed in clinical guidelines.
Is total body irradiation using a non-neutron producible beam safe and feasible for cardiac implantable electronic devices?
Is total body irradiation using a non-neutron producible beam safe and feasible for cardiac implantable electronic devices?
Total body irradiation is feasible and safe for patients with CIEDs if a non-neutron producible beam is used at a low dose-rate and shock therapy is deactivated.
Direct CIED irradiation in TBI may be feasible; hypothesis-generating and requires clinical validation before practice change.
Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in the literature. The feasibility of TBI while using a CIED was discussed for the first time.
No takes yet. Share an insight, caveat, or question.
Matsubara et al. (2021) studied Total body irradiation with cardiac implantable electronic devices (n=18). Direct irradiation by a 6-MV X-ray beam was evaluated on CIED malfunction. Direct irradiation of cardiac implantable electronic devices up to 20 Gy caused no malfunctions when shock therapy was deactivated, suggesting total body irradiation is feasible and safe.