This method demonstrates enhanced signal intensity for cholesterol in biosensors, suggesting improved disease detection.
The high and stable signal intensity of electrochemiluminescence (ECL) is crucial for establishing highly sensitive ECL biosensors. Here, nitrogen-vacancy/oxygen-doped carbon nitride (OCNNV-650) was first used as an ECL emitter, significantly improving the low ECL activity and poor dispersion of graphitic carbon nitride (g-C3N4) through a molecular engineering strategy. It is particularly noteworthy that Fe3O4@MoS2 peroxidases, as coreaction accelerators, are the key to achieving the first and last "signals on", promoting the production of more SO4•- and even promoting the production of •OH radicals when cholesterol is electrocatalyzed to produce H2O2. The intermediate "signal off" is induced by nitrogen and sulfur codoped carbon dots (NS-CDs), which significantly reduce the ECL detection background by hindering the electron transfer between SO42- and SO4•-. The optimized sensor detects 0.31 μM cholesterol with a broad detection range (0.1 μM-0.1 mM). The "on-off-on" strategy facilitates multipathway ECL emission of OCNNV-650 with high and stable signal intensity, presenting a new method to develop ECL biosensors with high sensitivity and real-time accurate detection for disease prevention and human health care.