The development of luminol-based electrochemiluminescence (ECL) biosensors has been constrained by the intrinsic instability of hydrogen peroxide and the sluggish activation of dissolved oxygen. We report a water-driven ECL platform based on cobalt-manganese nanosheets (CoMn-NS), which exhibit dual catalytic functionalities integrating interfacial oxygen evolution reaction (OER) electrocatalysis with oxidase-like nanozyme activity. In this system, O2 generated through OER is sequentially and efficiently converted into superoxide anion radical (O2·-) and singlet oxygen (1O2) via nanozyme catalysis, yielding a substantial 61-fold signal amplification in the co-reactant-free luminol/H2O system. A dual-quenching strategy integrating reactive oxygen species elimination with energy transfer was further established for target-responsive ECL signal modulation. Leveraging this design, the immunosensor achieved ultrasensitive detection of N-terminal pro-B-type natriuretic peptide with an impressive detection limit of 2.5 fg·mL−1 (linear range: 10 fg·mL−1 to 100 ng·mL−1) and demonstrating high recoveries (97.30% to 100.3%) in human serum. This ECL sensing strategy demonstrates excellent practical applicability and establishes an electrocatalytic−nanozyme paradigm for constructing stable water-driven ECL interfaces.