Low-cost system measures hydrocarbon gases with sensitivities of 111.6 ppm for isobutane, suggesting potential for gas sensing applications.
Spherical acoustic resonant cavities have been increasingly reported in photoacoustic spectroscopy due to their small volume and enhanced effective gas absorption path length. For further reducing the acoustic cavity volume and exploiting broadband LED as a light source, this paper reports a low-cost, LED-based photoacoustic gas-sensing system using a hemispherical acoustic resonant (HAR) cavity with a radius of 15 mm and a volume of 7.07 mL. The placement of both the excitation light source and transducer, as important elements in photoacoustic spectroscopy, was systematically optimized for improving the generation efficient of photoacoustic signal. The frequency response of the HAR cavity was thoroughly characterized for exploring an optimal operation frequency of the light source. Through positional and frequency optimization, the developed low-cost, LED-based photoacoustic spectrophone realized highly sensitive measurements of hydrocarbon gases with measurement sensitivities of 111.6 ppm (3σ) for isobutane, 140.1 ppm (3σ) for propane, and 866.4 ppm (3σ) for ethylene at an integration time of 1 s. These results demonstrate the strong potential of low-cost, LED-HAR-based PA-sensing systems in the field of gas sensing for widespread deployment in distributed sensor networks and atmospheric monitoring platforms.