Observational analysis showed this mattress accurately monitors heart rate and respiration rate, indicating its potential for daily health tracking.
A non-intrusive mattress based on an optical-fiber Michelson interferometer is designed for daily sleep monitoring. The optical phase signal of the optical-fiber Michelson interferometer caused by the heartbeat and respiration is demodulated by the phase-generated carrier (PGC) method. The physiological signals and vital indicators including heart rate (HR), respiration rate (RR), and signal energy (SE) are extracted from the optical phase by algorithmic processing. A series of experiments are conducted to confirm the feasibility of the mattress for sleep monitoring. The mattress not only can achieve HR and RR counting, but also can record the waveform of the sleep-induced signal accurately. The body states can also be distinguished by the SE. In an all-night sleep monitoring experiment, the HR measured by the mattress is compared with the HR measured by a commercial smart band, showing a maximum error of 6 bpm (beat per minute). The designed mattress based on an optical-fiber Michelson interferometer shows good performance and great potential in non-intrusive sleep monitoring.