Key result
The fiber optic smart mattress accurately monitored vital signs with maximum errors of 2 bpm for heart rate and 1 bpm for respiration rate compared to reference devices.
Why the study?
Can a smart mattress based on optical fiber Mach-Zender interferometer accurately monitor vital signs noninvasively?
Observational (n=18)
Can a smart mattress based on optical fiber Mach-Zender interferometer accurately monitor vital signs noninvasively?
Effect estimate: Max errors of HR 2 bpm and RR 1 bpm
A novel fiber optic smart mattress can accurately and noninvasively monitor heart and respiration rates, showing potential as a household medical device.
May support contactless monitoring feasibility; leaves open validation in clinical populations before practice adoption.
A smart mattress based on optical fiber Mach-Zender interferometer (OF-MZI) is designed for noninvasive and continuous monitoring of human vital signs. Through arranging the sensing fiber between two elastic covering layers with sandwich structure, the mattress was sensitive to the respiration and heartbeat induced micro-pressure. In the processing terminal, the waveforms of vital signs were demodulated by 3 * 3 coupler based differentiate and cross-multiplying method, and then four characteristic indicators including the heart rate, heartbeat amplitude, respiration rate, and respiration amplitude were respectively extracted through feature extraction algorithm, for evaluating the human health condition. Clinical experimental results of eighteen subjects indicate that the mattress system could not only distinguish the activity states of no body, on bed, body movement and off bed, but also contribute to clinical diagnosis of bradycardia, tachycardia, polypnea and apnoea. By adopting Bland-Altman analysis method, good reproducibility and accuracy were confirmed, where the max errors of heart rate and respiration rate are respectively 2 bpm and 1 bpm. Moreover, the responses at different positions of the mattress are identical and the continuous monitoring results in one day are consistent with daily change of vital signs, which proves that the fiber optic mattress has good reliability and stability. Beneficial from high-sensitivity, multiple parameters, long-term continuous monitoring, high comfortability and low cost, the mattress is promising in the early detection and prevention of cardiac and respiratory diseases as a household medical device.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2020) conducted an observational in Cardiac and respiratory monitoring (n=18). Fiber optic Mach-zender interferometer (OF-MZI) smart mattress vs. Commercial oximeter and direct observation was evaluated on Accuracy and reproducibility of heart rate and respiration rate measurements (Max errors of HR 2 bpm and RR 1 bpm). The fiber optic smart mattress accurately monitored vital signs with maximum errors of 2 bpm for heart rate and 1 bpm for respiration rate compared to reference devices.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: