Observational study compares manual and electronic blood pressure recordings in ICU patients, indicating discrepancies in accuracy.
This cross-sectional observational study aimed to evaluate the difference between manual and electronic blood pressure (BP) recordings among 112 adult ICU patients in a tertiary care hospital. Systolic and diastolic BP readings were taken using both manual auscultatory and electronic oscillometric methods during morning and evening hours. Results revealed that manual readings consistently overestimated BP, with a mean difference of 7.90 mm Hg for systolic and 7.77 mm Hg for diastolic values compared to electronic measurements. Over 97% of patients had a ≥5 mm Hg difference, and nearly 74% had a ≥10 mm Hg difference. The overestimation was uniform across gender and age groups, though slightly more pronounced in older adults. Minimal diurnal variation was observed. Cross-classification analysis showed that 44% of patients were misclassified into higher BP categories by manual readings, potentially impacting clinical decisionmaking. Despite strong correlation coefficients between the two methods, the consistent bias suggests that electronic monitoring offers more reliable data in critical care. These findings highlight the need to reevaluate the reliance on manual methods in ICU settings, especially in cases requiring precise BP control for medication titration and risk stratification.