Intraoperative sensing tool improved stiffness assessment in tissue samples, suggesting biocompatibility and safety.
A novel endoscopic palpation tool (EPT), designed for tactile and stiffness sensing using fiber Bragg gratings (FBGs) was evaluated in a surgical environment for intraoperative safety and effectiveness. The EPT consisted of four FBGs arranged in a cross pattern and embedded within an elastic, hollow, silicone hemispherical dome designed to deform upon contact with soft tissue. The EPT was employed to scan both in vivo and ex vivo tissue samples. Monitoring of porcine vital signs during minimally invasive and open surgical procedures showed no significant changes during use of the EPT. Perioperative blood tests including inflammatory markers and liver and renal function studies were unremarkable. Histopathological analyses of tissues involved (liver, spleen, bowel, and abdominal wall) showed no evidence of inflammation, necrosis, or tissue damage, confirming the device’s biocompatibility. To the best of our knowledge, this is the first study reporting in vivo stiffness measurements using an FBG-based EPT. The probe successfully distinguished between soft and hard tissue regions’ relative stiffness. Furthermore, successive measurements on liver samples demonstrated the device’s ability to generate stiffness maps, enabling clear visualization of spatial variation in tissue stiffness.
No takes yet. Share an insight, caveat, or question.
Violakis et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: