Retrospective study compares black blood and bright blood T2 techniques for myocardial iron load in thalassemia, suggesting interchangeability.
Objective To investigate the difference of black blood (DB) and bright blood (BB) T2* techniques at 1.5 T and 3 T in the assessment of myocardial iron load in patients with thalassemia (TM). Methods As a retrospective study. CMRtools software was used to measure myocardial T2* in 359 patients with moderate (60 g/L < Hemoglobin<90 g/L) or severe (Hemoglobin<60 g/L) thalassemia. A truncation method was used to remove signal values that deviated from the fitted curve. T2* (DBx-T2*, BBx-T2*) containing all (eight echoes) signals (DB8-T2*, BB8-T2*) and the optimal signal (coefficient of determination R 2 > 0.95) were recorded. The difference, correlation and consistency of T2* measured by different methods were compared. Results There was no significant difference ( p > 0.05) in myocardial T2* measured by different methods (1.5 T, 3 T), and all of them were highly positively correlated ( p < 0.05, r s > 0.9). Bland–Altman analysis showed that (1.5 T) DB8-T2* and DBx-T2*, DBx-T2* and BBx-T2* had good consistency ( p > 0.05). (3 T) DB8-T2* and DBx-T2* had good consistency ( p > 0.05). There were proportional differences in T2* values measured by the other methods ( p < 0.05), and none of them could be considered to have good consistency. Conclusion DB CMR T2* and BB CMR T2* can be interchangeable in the assessment of myocardial iron load in TM patients. However, DB CMR T2* is more stable and reliable.