Observational analysis reveals significant polymerization delay in resin-based composites, indicating thickness and irradiance both influence the curing process.
To examine the effects of the resin-based composite (RBC) thickness and incident irradiance on the time delay after the light turns on before polymerization starts at the bottom surface of the RBC. The degree of conversion (DC) was determined using Fourier-transform infrared spectroscopy (FT-IR) at a rate of 13 measurements per second. Six commercial RBCs were tested at three thicknesses (0.2, 2 and 4 mm) and four incident irradiance values (472-3985 mW/cm2). There was a significant time delay before polymerization began at the bottom of each RBC. At 942 mW/cm2, representative of many current LCUs, this delay ranged from 0.12 to 1.11 s at the bottom of 2 mm of RBC. When the thickness was increased to 4 mm, even with an irradiance of 1969 mW/cm2, which represents high-power LCUs, the delay ranged from 0.12 to 5.17 s. The highest irradiance of 3985 mW/cm2 did not fully overcome the inherent lag caused by the thickness of the RBC. At 3985 mW/cm2, PowerFlow and PowerFill had the shortest delay times of 0.09 s and 0.30 s, respectively, at the bottom of a 4 mm thick RBC. The presence of a time delay before photo-cured RBCs start to polymerize at the bottom surfaces must now be addressed clinically and when calculating reaction rates.
No takes yet. Share an insight, caveat, or question.
Price et al. (2025) studied this question.