Radiant exposures delivered by the various irradiation protocols were between 4.5-30.3 J/cm². Short exposure times from MONET-1 s and PW-3 s delivered the lowest radiant exposures (4.5 and 5.2 J/cm², respectively) and produced the lowest ΔT and DC. The longer exposure times in the standard modes of PW-20 s, S10-20 s, and MONET-3 s produced the highest Tmax, ΔT, and DC for both composites. The ΔT range among composites at different depths varied significantly (31.7-49.9 °C). DC of TPF ranged between 30-65% and in EVO between 15.3-56%. TPF had higher Tmax, ΔT for all depths and DC compared to EVO, across the LCU protocols (p < 0.05), except for PW-20 s and MONET-3 s. The coronal part of the restorations (1-2 mm) had the highest ΔT. There was a positive correlation between ΔT and DC at 4-mm depth after 120 s SIGNIFICANCE: Longer, or standard, exposure times of the LCUs delivered greater radiant exposures and had higher DC and ΔT compared to shorter or high-irradiance protocols. The fast photo-polymerised RBC had comparatively superior thermal and conversion outcomes when it received a high irradiance for a short time (1-5 s) compared to the conventional Bulk-Fill RBC.