Simulation assessed tidal volume and pressures in two lung simulators, indicating risks in simultaneous ventilation.
Background The concept of ventilating multiple patients concurrently using a single ventilator has been proposed as a solution when the demand for ventilators surpasses the available supply. While the practicality of this approach has been established, a thorough evaluation of the risks involved has yet to be comprehensively addressed. Methods Two circuits, a simple one (circuit-1) and another with an adjustable resistance valve (circuit-2), were evaluated within an experimental framework utilizing two computer-controlled lung simulators (TestChest and ASL 5000). These simulators were ventilated by an ICU (intensive care unit) ventilator (Servo-u) employing various ventilation modes (volume- and pressure-controlled ventilation). The study was conducted under differing respiratory conditions, characterized by low compliance (20 ml/cmH 2 O) as well as normal-high compliance (100 ml/cmH 2 O), in order to ascertain the applied tidal volume (VT), pressures, and the resultant mechanical power (MP). Results Circuit-1: The applied VT, pressures, and MP differed significantly between the two simulators, as well as in relation to ventilation mode, compliance, and respiratory rate (RR) (p < 0.001); the differences were most pronounced in settings with differing compliance levels. Circuit-2: Differences in VT, pressures, and MP were observed between simulators concerning valve settings (p < 0.001). The VT demonstrated a negative correlation, with volumes derived from valve closure spanning from 50 to 100 ml across all settings. In the design of circuit-2, MP exceeded the 12 J/min threshold in both lung simulators at elevated RR and could only be decreased through valve closure followed by a consequential hypoventilation in one simulator. Conclusion The simultaneous ventilation of two patients using a single ventilator is technically viable, yet it presents considerable risks. Even with the integration of an adjustable resistance valve to accommodate varying lung complexit i es, the likelihood of unilateral hypoventilation and elevated mechanical stress remains high.
No takes yet. Share an insight, caveat, or question.
Harnisch et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: