Answers are grounded in this paper and cite their sources.
Evaluation reveals PRFS thermometry can accurately monitor temperatures in homogeneous phantoms, suggesting potential use in interventional MRI safety.
Motivation: Evaluate the feasibility of proton resonance frequency shift (PRFS) thermometry at 0.55T while assessing achievable accuracy and its potential use in interventional MRI ablation procedures and safety evaluations. Goal(s): Determine the performance of PRFS thermometry at 0.55T against calibrated fiber-optic temperature probes. Approach: Perform phase-shift, drift-corrected thermometry on a homogeneous agar/copper sulfate phantom during heating and cooling with simultaneous temperature probe measurements. Results: Temperature change was measured for up to one hour with drift correction, and measurements correlated well with the fiber-optic probes. The mean difference between measurements was 0.93±0.35°C and average limit of agreement window (±1.96 SD) was 3.86°C. Impact: This study explored the accuracy of thermometry at 0.55T in simple homogenous phantoms while looking ahead to implementation in iMRI applications. Thermometry at 0.55T has potential for real-time temperature monitoring with a clinically relevant degree of accuracy.
No takes yet. Share an insight, caveat, or question.
Schepers et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: