Low field strength MRI scanners are appealing for MR guided interventions because of their lower associated costs and increased compatibility with interventional devices. Cardiac PRFS thermometry shows promise for real-time guidance of ablation therapy of cardiac arrhythmias. In this study, initial characterization of cardiac PRFS thermometry at 0.55T was performed in 5 healthy volunteers by measuring the stability over time. The stability of thermometry was 1.6±0.8 °C in the myocardium, confirming feasibility at this field strength.
No takes yet. Share an insight, caveat, or question.
Mooiweer et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: