Optimization framework reduces SAR in images of metal orthopedic hardware, suggesting significant scan time savings.
Motivation: Metal artifact reduction sequences suffer from SAR resulting in longer scans. Goal(s): Our goal was to reduce the sequence SAR without compromising image contrast and image quality. Approach: We propose an optimization framework to design variable refocusing flip angles to minimize SAR while maintaining image quality comparable to clinical sequence. The prototype sequence was tested on a cohort of patients with metallic implants at 1.5T. Results: The variable flip angle sequence generates images of comparable quality as conventional clinical acquisition while allowing savings in SAR and acquisition time. Impact: Metal artifact reduction sequences that utilize the proposed variable flip angle framework result in lower SAR and scan time without any loss in image quality.
No takes yet. Share an insight, caveat, or question.
Keerthivasan et al. (2025) studied this question.