Phantom evaluation demonstrates feasible spectrally-encoded multi-spectral imaging at 0.55T, indicating improved imaging adjacent to metal implants.
Spectrally-encoded multi-spectral imaging (SEMSI) enables imaging adjacent to metal implants without blurring due to off-resonance, and has been demonstrated at 3T. Here, we demonstrate application of this technique at 0.55T with appropriate adjustments to the spectral resolution and RF pulse bandwidth. A total hip arthroplasty phantom was imaged with SEMSI, turbo spin echo (TSE), and Slice Encoding for Metal Artifact Correction (SEMAC). Experiments demonstrate the feasibility of SEMSI at 0.55T, and examine ΔTE to support multi-echo acquisitions in a single TR.