Metallic implants cause severe distortions in the magnetic field that are best mitigated using multi-spectral acquisitions that require longer scan time. At conventional field strengths, this is compensated by aggressive use of parallel imaging. However, at low-field strengths (<1.5T), available parallel imaging factors are reduced because body noise dominance prevents the use of dense arrays with smaller elements. Hexagonal sampling (in ky,kz space) was previously proposed as a way to achieve 2-fold reduction in imaging time without introducing additional artifacts. In this work, we demonstrate applicability of the hexagonal sampling for 0.55T multi-spectral imaging near metal.