Electromagnetic simulations reveal coil element deactivation increases peak sar by 77.61%, highlighting safety implications.
Motivation: In 7T cardiac MRI, controlling peak SAR and SAR distribution is critical for safety. This study explores how deactivating loop coil elements affects these metrics. Goal(s): We assess the impact of coil element deactivation on SAR in a shared-conductor loop coil across different human models. Approach: Electromagnetic simulations were conducted on ten human models, analyzing SAR under normal conditions and with individual coil element deactivations. Results: Coil element deactivation increases peak SAR, up to 77.61% in the Rosalind model, and changes SAR distribution by up to 474% in the Laura model, varying with coil position and model anatomy. Impact: This study highlights the significant impact of coil element deactivation on SAR, requiring a safety factor in clinical routines to protect patients if coil elements become deactivated.
No takes yet. Share an insight, caveat, or question.
Jabbarigargari et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: